Products - Zhengzhou Smile Dental Equipment Co., Ltd
Home > products
     - DP32
    DP32

    Integral Type Suction

     

    DP32 Integral Type Suction

    Integral type suction

    Noise reduction,low vibration

    Doors can be completely disassembled,easy to install

    Separator and drainage bottle can be removed at any time to clean and install

     

    DP32 DS3701M

     

    Integral Type Suction

     
     - SS11
    SS11

    18/23 Liters Autoclave Sterilizer
    Class B

     

    Specifications:

    Model
     
    JY-18
    JY-23
    Capacity
     
    18L
    23L
    Sterilizing temperature
    121℃,134℃
    Special function
    kill the  AIDS(HIV),hepatitis B(HBV)mad cow virus and Bacillus
    Dry procedure
    vacuum drying
    Display
    digital LED display with high definition
    Sterilization program
    4 kinds of programs
    Test program
    Vacuum test,Helix test
    Sterilizing data
     
    print by optional external mini-printer or download by USB port
    Safety features
    Safe Valve;Double-control lock system and self-test alam system
    Water supply system
    two water tanks inside,capacity of reservoirs(3.5L fresh water and 5L waste water)
    AC power
    220V/50Hz
    Rated power
    1600W
    Instrument containers
    5-layer activity pallet rack with 3 trays
    Chamber size
     
    250X355mm
    250X450mm
    N.W./G.W.
     
    45/48KGS
    47/50KGS
    Packing size
     
    660X550X560mm
    660X550X560mm

     

     Technical Features:

    1.Steam generator and heating coil work separately.
    Separate heating work of steam generator and heating coil lead to easy manipulation of the speed and volume of steam, reducing the water consumption. Heating coil lifetime is prolonged because its temperature will fall when chamber temperature goes up.
    2.Material of cover is safe and reliable
    Covers and chambers are made of imported 2mm SU304 stainless steel which will guarantee safety at a high pressure inside chambers and avoid the second pollution caused by the corrosion.
    3.Advantages of the control system
    Getidy innovates and develops new technology to control temperature and pressure and ensure the tolerance temperature inside chamber is less than 0.5℃.
    4.The reliability of measurement accuracy
    The accuracy of temperature is 1℃,and the accuracy of pressure is 0.1bar.
    5.The safety of double-control lock
    Use the newest machinery and electronic double-control lock ,to ensure the safety during the operation.
    6. Automatic and manual adjustment of temperature sensor
    The temperature sensor can adjust automatically or manually at will to ensure the autoclave work in different atmospheric environments.
    7. Multilanguage are optional, Easy to operate
    Users can easily find the language according to their requests.
    8.Broad ventlight makes good heat dissipation

    Instructions:

    1. Please read the manual carefully before operation.
    2. Put sterilizers on horizontal worktable and raise the front two supporting stands slightly.
    3. Be sure to use distilled water in order to prolong the life of sterilizers.
    4. Clean the water tanks and water filter regularly.
    5. Do not move the sterilizer during the process of work.
    6. All the medical equipment, blankets, clothing etc. which contact pathogenic microorganisms must be sterilized by chemical disinfectant first.
    7. The package or container for disinfection items must be appropriate, and is recommended to use double wrapped white cotton with good steam penetration.
    8. Try to put the same kind of instrument in the same sterilizer flatly,leave aperture, make sure steam can penetrate easily.
    9. Make sure the machine in good condition, especially the safe valve.
    10. Do not open the door until the pressure fall down to “0” bar.

     

     
Total 1535 Items 768 pages/Current is Page 87     First | Last | Previous | Next    go topage